Thermowell system HIT: Hygienic invasive temperature measurement

Type series HP1100

Thermowell system HIT: Hygienic invasive temperature measurement - HP1100

Features

  • Hygienic invasive thermowell system
  • Installation and disassembly without interruption of process
  • Wetted parts stainless steel mat.- no. 1.4435 (316L)
  • Surface roughness ≤ 0,4 μm, electropolished
  • Deltaferrite content < 3%
  • Material certificate per DIN EN 10204-3.1
  • Pipe standards: DIN 11866 series A / DIN 11850; series B / ISO1127, series C / ASME BPE
  • Design: straight or angled pipe system
  • Instrument connection M12 × 1
  • Nominal pressure: up to PN 25
  • Suitable temperature sensor MiniTherm GA 2730

Options

  • Deltaferrite content < 1 %

Technical Data

Constructional design / case
Design Thermowell, orbitally welded into a pipe system
- Straight
- Angled (90 °)
Instrument connection M12 x 1
Nominal width See data sheet
Nominal pressure Up to PN 25 per DIN 11865
Weight See data sheet
Material wetted parts
Thermowell Stainless steel mat.-no. 1.4435 (316L)
Pipe system Stainless steel mat.-no. 1.4435 (316L)
Deltaferrite content < 3 %, optional < 1 %
Surface wetted parts
Surface treatment Electropolished
Roughness (inside) ≤ 0.4 μm (except welded seam)
Temperature sensor
Temperature sensor See GA2730, data sheet T4-017
Screwing Spring loaded union nut M12 x 1 with spring contact system for best heat transfer
Insertion length Circular connector l1 = 29 mm
Field housing l1 = 33 mm
Design without neck tube
Accuracy without heat sink compound t90 = 7 s
with heat sink compound t90 = 6 s
Type series
HP1100

Documents

Name
Description
EN
DE
  • TA_033

    Technical Information - EX zones separation

  • T5-050

    Data Sheet

Productrequest