Constructional design
Electrical connection: | Circular connector M12 (4 pin) Option: Circular connector M12 (8 pin) for 2 x Pt100, Field housing Ø 60 mm with screw cap, rotatable, positionable through ± 170° Material: stainless steel mat.-no. 1.4305 (303) |
With cable glands: - M12 x 1,5 PA black - M12 x 1,5 stainless steel - M16 x 1,5 PA black |
|
Weight: | With circular connector M12: - pipe-Ø ≤ 17.2 mm: approx. 100 g - pipe-Ø ≥ 18.0 mm: approx. 200 g |
With field housing: approx. 400 g | |
With transmitter integrated in the circular connector M12: - pipe-Ø ≤ 17.2 mm: approx. 130 g - pipe-Ø ≥ 18.0 mm: approx. 230 g |
|
Type plate: | Adhesive label |
Measuring insert
Design: | Special measuring insert: Ø 6 mm; hygienic design. Measuring insert screwed into the clamp-ing element under spring tension. |
Material: | Stainless steel Measuring element from silver, thermally isolated via PEEK element. |
Measuring resistor: | - Pt100 per EN 60751, class A 4-wire (also connectable in 3-wire) |
- Pt100 per EN 60751, class A 4-wire (3-wire bridged) | |
- Pt100 per EN 60751, class A 3-wire | |
- 2 x Pt100 per EN 60751, class A 3-wire | |
Degree of protection per EN 60529: | IP 67 |
Process connection
Design: | Clamping element designed for installation with: - clamping block for pipes Ø 4...57 mm - clamping shoe for pipes Ø 10...300 mm |
Material: | Temperature resistant plastics (PVDF) with integrated isolation system, hygienic design |
Degree of protection per 60529: | IP 65 |
Pipe diameter: | See order code |
Accuracy
The accuracy and response time of the whole system depend on the pipe geometry, the medium and the ambient temperature. |
For Pt100 per EN 60751, class A 4-wire (also connectable in 3-wire): Accuracy of system in the range -20 up to 150 °C: For all nominal ranges: (TU - TM) x 0.02 * |
For 2 x Pt100 per EN 60751, class A 3-wire): Accuracy of system in the range -20 up to 150 °C: For all nominal ranges: (TU - TM) x 0.035 * |
For Pt100 per EN 60751, class A 3-wire and 4-wire (3-wire bridged): Accuracy of system in the range -20 up to 150 °C: - For nom. ranges ≥ 18.0 mm: (TU - TM) x 0.02 * - For nom. ranges < 18.0 mm: (TU - TM) x 0.03 * (TM = media temperature, TU = ambient temperature) |
Repeatability: typical 0.1 °C, max. 0.2 °C * |
Response time: t90 = 8...15 s * (on pipe Ø 18 x 1.5) |
* with use of heat sink compound (see Type MT8800) |
Temperature ranges
Ambient: | -20...80 °C |
Storage: | -40...80 °C |
Transmitter
Installation variants: | - Transmitter, Type PA2430, for circular connector M12 |
- Transmitter, Type PA2530 IO-Link, for circular connector M12 | |
- Transmitter head mounted, Type series PA210., 4…20 mA, programmable | |
- Transmitter head mounted, Type series PA220., electrically isolated, classification per SIL 2 | |
- Transmitter head mounted, Type series PA230., electrically isolated, classification per SIL 2, HART® | |
- Transmitter head mounted, Type series PA2420, 2 channel, classification per SIL 2/3, HART® |
Tests and certificates
see documents |
Type series
GA2610, GA2611 |